Contact an Account ManagerPCB from waste oil
Summary
Type: SPE
Application #: 301390
Catalog#: 730132 - SPE Cartridges, Chromabond, SA/Silica, 3ml, PP, 500/500mg 50/pk
Detection:
Conditions: Column conditioning: 1 ml n-hexane
Sample application:
Apply about 250 µl waste oil sample to the column and aspirate or force it into
the adsorbent with 2 x 1 ml n-hexane
Elution:
Aspirate o
Authors: Macherey-Nagel
Source: LC Department
Details
Legend: Compound recovery [%]
PCB 28 97
PCB 52 96
PCB 101 95
PCB 138 90
PCB 153 95
PCB 180 96
PCB 209 100